EasyManua.ls Logo

Philips IntelliVue MX40 - MX40 Performance at End of Battery Life

Philips IntelliVue MX40
184 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Monitoring Mode
Operating Modes 4-27
MX40 Performance at End of Battery Life
Note 1 – If the SRR is already connected, SRR connectivity continues. In this
state, it is not possible to start a new SRR connection.
Note 2 – If the display is already turned on, the display will dim, but will
continue to function. In this state, it is not possible to turn the display on if
it is off.
Note 3—If the Tele Battery Low INOP occurs and then the MX40
disconnects from the network, the behavior and timing described in the
diagram above does not apply. The screen activates to show a large red
battery icon , and the device plays an escalating double beep sound to
indicate that it is disconnected from the network. Alarm and SpO2
monitoring are no longer available locally. The MX40 will attempt to
reconnect to the Information Center to transmit the battery INOP condition.