Cleaning
Maintenance 5-3
1. Remove the batteries and disconnect the patient cable. Note that
disconnecting the patient cable for cleaning is dependent on your
hospital’s protocol. The connection between the MX40 and the
patient cable is rated IPX7 (protected against the effects of
temporary immersion in water). Care must be taken to ensure this
area of connection is completely dry prior to reconnecting the MX40
with the patient cable. The MX40 and patient cable must be
connected correctly and completely in order to maintain the IPX7
rating. The battery compartment of the MX40 has an IPX3 rating
(protected against the effects of spraying water).
If using disposable AA batteries, remove the battery adapter tray
and clean separately. Dispose of AA batteries according to hospital
policy.
If using the MX40 rechargeable battery, remove and clean
separately.
2. Clean the MX40, rechargeable battery, patient cables, and battery
adapter tray before disinfecting.
3. Wipe the MX40, rechargeable battery, and patient cables using a
lint-free cloth dampened modestly with one of the approved
cleaning or disinfecting agents listed in the table below.
If using a cotton swab to clean or disinfect, swabbing should be
unidirectional at a time. Parallel strokes should be used to cover the
entire swab area as shown below.
Start
End
Start
4. Wipe the exterior surfaces of the patient cable using a lint-free cloth
dampened modestly with one of the approved cleaning or disinfection
agents listed in the table below. If the interior surfaces of the patient
cable connector require disinfection, wipe using a lint-free cloth
dampened modestly with isopropyl alcohol only. Wipe between and
around the MX40 connector pins to remove chemical residue.
5. Wipe the outside surfaces of the battery adapter tray with one of the
approved cleaning or disinfecting agents listed in the table below.
Wipe the inside surfaces of the battery adapter tray with a lint free
cloth dampened modestly with isopropyl alcohol.
6. Follow the manufacturer's instructions with regard to application
duration.