MX40 Controls in the Patient Window (IIC iX)
Monitoring with the MX40 at the Information Center 11-5
MX40 Controls in the Patient Window (IIC iX)
The Patient Window at the Information Center iX includes controls for a
number of MX40 operations. For detailed instructions on these operations,
see the IntelliVue Information Center iX Instructions for Use or the Online Help.
To View ECG, Resp, NBP or SpO
2
Alarm Limits
1 Click on the HR or SpO
2
label in the Patient Window to display the
current high and low alarm limits.
To Change ECG, Resp, NBP or SpO
2
Alarm Limits
1 Click on the HR or SpO
2
label in the Patient Window.
2 Click on High Limit or Low Limit and select the new value from the list.
To Change Paced Status
1 Click on the HR label in the Patient Window.
2 Click on Paced Mode.
3 Click on Off or On to select the status.
Warning
At admission/discharge, always check that paced status is correct for the
patient.
To Change ECG Waveform Size
1 Click on the appropriate waveform.
2 Select Size Up or Size Down.
To Select Lead
1 Click the appropriate waveform.
2 Select the desired lead label from the list.
Important Do not set the primary and secondary channels to the same
lead.
To Change Va and Vb Default Lead Settings (6-lead
only)
1 Click the Measurements Application button.
Draft Copy