Electromagnetic Compatibility
Safety Standards & Specifications 15-13
Electrosurgery Interference/Defibrillation/Electrostatic
Discharge
The equipment returns to the previous operating mode within 10 seconds
without loss of any stored data. Measurement accuracy may be temporarily
decreased while performing defibrillation. This does not affect patient or
equipment safety. Do not expose the equipment to x-ray or strong magnetic
fields (MRI). The MX40 is not for use during electrosurgery.
Restart Time
After power interruption, an ECG wave will be shown on the display after
30 seconds maximum.