Environmental Specifications
Safety Standards & Specifications 15-25
Environmental Specifications
Parameter
Specification
Temperature
Operating
Storage
0 to 37
o
C (32 to 99
o
F)
-30
o
C to 50
o
C (-22
o
F to 122
o
F) without batteries
12
o
C to 35
o
C (53.6
o
F to 95
o
F) with Single-Patient-Use
leadsets
Humidity
Operating
Storage
< 95% RH at 37
o
C (98.6
o
F) non-condensing
< 90% RH at 50
o
C (122
o
F) without batteries
Altitude
Operating &
Non-operating
Barometric
Pressure
3,000 m (9,842 ft)
72kPa (537 mmHg)