Default Settings B-3
ECG, Arrhythmia, ST and QT Default Settings
ECG Settings
Factory Defaults
Adult
Pedi
ECG
On
Primary Lead
II
Secondary Lead
6-lead: III
5-lead (Standard): III
5-lead (EASI): III
Default ECG Size
x1
Lead Placement
Standard
Leadset Type
AAMI
Analysis Mode
Multi-lead
High Limit
120 bpm
160 bpm
Low Limit
50 bpm
75 bpm
Asystole Threshold
4.0 sec
Arrhythmia Settings
Factory Defaults
Adult
Pedi
Arrhythmia
On
Pause Threshold
2.0 sec
VTach HR
100 bpm
120 bpm
VTach Run
5
Vent Rhythm
14
SVT HR
180 bpm
200
SVT Run
5
PVCs/min
10
5
Non-Sustained VT
On
Run PVCs
On
Pair PVCs
On
R-On-T PVCs
On