19 Assigning a Telemetry Device and a Monitor to One Patient
267
In case of ambiguity, a yellow INOP message !!CHECK ECG SOURCE indicates that more than one
valid ECG source is active.
Synchronized Settings
If ECG is measured at the monitor, and then the patient is connected to a telemetry device for
monitoring, the Information Center will use the monitor settings for the telemetry device. In general,
the following settings will be synchronized:
WARNING
• Not all settings are synchronized; after changing the ECG source, always check that the settings
are appropriate.
• Va and Vb leads are reset to default (V2, V5) if the configured Va or Vb lead for the telemetry
device is not one of V1 through V6
If later the patient is disconnected from the telemetry device, and reconnected to the monitor again,
any changes in the settings made in the meantime will be passed on to the monitor. In this way,
settings continuity is preserved when the ECG source changes.
Settings synchronization can only take place when there is no patient information mismatch between
the monitor and the Information Center. If a
Check ECG Settings or !! CHECK PAIRING INOP
appears always check that the ECG settings, especially the paced setting, are appropriate for your
patient.
Heart Rate HR/Pulse Alarm On/Off, Heart Rate High/Low Limit,
ECG Primary Lead, Secondary Lead, Va Lead
1
, Vb Lead
1
, Lead Placement
Arrhythmia Analysis Mode, Asystole Threshold, Pause Threshold, VTach HR, VTach
Run, PVCs/min, Vent. Rhythm, SVT HR, SVT Run, PVCs/min On/Off,
Pacer not capture On/Off, Pacer not pace On/Off, Non-Sustain On/Off,
Vent. Rhythm On/Off, Run PVCs On/Off, Pair PVCs On/Off, Missed
Beat On/Off, Pause On/Off, R-on-T PVCs On/Off, Vent. Bigeminy On/
Off, Vent. Trigeminy On/Off, Multiform PVCs On/Off, Irregular HR
On/Off, SVT On/Off, Afib On/Off
ST ST Analysis On/Off, ST Alarm On/Off, ISO Point, J point, ST point, ST
Priority List, Single ST Alarm Limit, Multi ST Alarm Limit
QT QT analysis On/Off, QT Lead, QTc High Alarm On/Off, ΔQTc Alarm
On/Off, QTc High Limit, ΔQTc High Limit, QTc Formula
2
, QT Baseline
SpO
2
TSpO
2
Alarms on/off, SpO
2
Alarm limits
NBP Alarm Suppression On/Off, Pulse(SpO
2
) On/Off
1
Va and Vb leads are reset to default (V2, V5) if the configured Va or Vb lead for the telemetry device is not one of V1
through V6.
2
This setting will only be synchronized when the Information Center supports QT Analysis for Telemetry