3 Patient Alarms and INOPs
70
Meas. DEACTIVATED An X2 or MP5 has been connected to a host monitor (companion mode) and all derived
measurements have been deactivated and/or measurements with a label conflict. The measurements
can only be reactivated by disconnecting the measurement device from the host monitor.
!!MORE BED ALARMS
!!!MORE BED ALARMS
at Information Center
The monitor is associated with a telemetry device and is sending data to the Information Center via
the telemetry device. There are currently more alarms at the bedside than can be transmitted to the
Information Center.
MSL Power High The power consumption of the devices connected to the Measurement Link (MSL) cable is too high.
If this situation continues, the MSL will be switched off. Contact your service personnel.
MSL Power Off
INOP tone
The power consumption of the devices connected to the Measurement Link (MSL) cable was too high
for too long and the MSL has been switched off. Contact your service personnel.
MSL Power Overload
INOP tone
The power consumption of the devices connected to the Measurement Link (MSL) cable is much too
high or there has been a short circuit. The MSL has been switched off. Contact your service personnel.
No Central Monit.
INOP tone
There is a problem with the communication to the network. Central monitoring is currently not
possible (no patient alarms or information). Check the connection. In case the connection is via a
telemetry device, the current telemetry use model does not support central monitoring. Contact your
service personnel.
NO ECG AT CENTRAL The ECG measured with the monitor ECG is not being sent to the Information Center via the
telemetry device.
Rem.AlarmDev.Malf.
INOP tone
There is a problem with the connection to the remote alert device. Contact your service personnel to
check the remote alert device and its connections.
Settings Malfunc.
INOP tone
The monitor cannot use the predefined settings for monitoring. Contact your service personnel.
Speaker Malfunct. Contact your service personnel to check the speaker and the connection to the speaker.
SRR INTERFERENCE
INOP tone
The short range radio connection has interference from another device. Try using another channel.
SRR INVALID CHAN
INOP tone
The channel configuration of the Short Range Radio is invalid. Check channel and channel mask
configuration.
SRR MALFUNCTION Malfunction in the short range radio device. If the INOP persists contact your service personnel.
TAAP DISABLED
INOP tone
The currently selected telemetry configuration on the monitor does not allow connection of telemetry
devices to the monitor.
TELE CONFIG UNSUPP
INOP tone
Telemetry device not supported (companion mode)
TELE EQUIP MALF
INOP tone
The telemetry device has a malfunction. Disconnect and reconnect the telemetry device. If the INOP
reappears, replace the telemetry device.
TELE INCOMPATIBLE SRR-enabled telemetry device is not supported by this central software revision. Please check
configuration.
TELE UNSUPPORTED
INOP tone
This telemetry device is not supported for direct connection to the monitor.
TimeExpired:<Timer
Label>
INOP tone
The time has expired for the timer indicated in the INOP text. Clearing the timer clears the INOP.
Unsupported LAN
INOP tone
There is a problem with the communication to the network and central monitoring is currently not
possible. Check the connection. If the INOP persists, switch off the monitor and contact your service
personnel.
User I/F Malfunct.
INOP tone
Perform a visual and functional check of all the monitor input devices. Contact your service personnel.
INOP Message, Indication What to do