Step 3. Set Up Your Service PC.
IntelliVue Telemetry System Infrastructure Installation and Service Guide 3-7
Step 3. Set Up Your Service PC.
To prepare your service PC to configure the APCs, you must:
Configure your PC to connect to the ITS Wireless Subnet.
Copy the Philips Upgrade Wizard tool and the latest ITS APC and AP firmware
files to your PC from the Philips Network Infrastructure Tools CD-ROM.
Configuring Your PC to Connect to the ITS Wireless Subnet
To configure your PC to connect to the ITS wireless subnet:
1. Configure your Service PC’s network adapter for 100 Mbs full duplex
communications and assign a fixed IP address to your Service PC that falls
within the ITS subnet address space:
a) Disconnect all network connections from your PC.
b) Select Start/Settings/Network Connections/Local Area Connection from
the Windows desktop.
The Local Area Connection Status dialog appears.
c) Click Properties.
Figure 5: ITS PoE Unit to ITS Synchronization Unit Cable Connections
ICN Network Switch
ITS Power over Ethernet Unit
Data & Power Ports (To APs)
Data Ports (To ICN Network Switch)
ITS Synchronization Unit
- From Power Hub (PoE Unit)
- From Access Points
Master/Slave
12345 6
12345 6
12345 6
12345 6
1
24
2