Measurement Specifications 24 Specifications
265
Invasive Pressure and Pulse
Complies with IEC 60601-2-34:2000/EN60601-2-34:2000.
Invasive Pressure Performance Specifications
Measurement Range –40 to 360 mmHg
Pulse Rate Range 25 to 350 bpm
Accuracy ±1% Full Range
Resolution
1 bpm
Input Sensitivity Sensitivity:5 µV/V/mmHg (37.5 µV/V/kPa)
Adjustment range:±10%
Transducer Load Impedance:200 to 2000 (resistive)
Output Impedance:3000 (resistive)
Frequency Response dc to 12.5 Hz or 40 Hz
Zero Adjustment Range: ±200 mmHg (±26 kPa)
Accuracy
±1 mmHg (±0.1 kPa)
Drift Less than 0.1 mmHg/°C (0.013 kPa/°C)
Gain Accuracy Accuracy ±1%
Drift
Less than 0.05%/°C
Non linearity
and Hysteresis
Error of 0.4% FS (@CAL 200 mmHg)
Overall Accuracy (including
transducer)
±4% of reading or ±4 mmHg (±0.5 kPa), whichever is greater
Volume di sp lacement of CPJ840J6 0.1 mm
3
/100 mmHg
Invasive Pressure Alarm Specifications Range Adjustment Delay
Pressure –40 to 360 mmHg
(–5.0 to 48 kPa)
-40 to 50 mmHg
2mmHg (0.5kPa)
>50 mmHg
5mmHg (1kPa)
max.
12 seconds
Extreme High Difference to high limit 0 to
25 mmHg
5mmHg steps (0.5kPa)
Clamping at -40 to
360 mmHg
5mmHg steps (1.0kPa)
Extreme Low Difference to low limit 0 to
25 mmHg
5mmHg steps (0.5kPa)
Clamping at -40 to
360 mmHg
5mmHg steps (1.0kPa)
Pulse 25 to 300 bpm Adult:
1 bpm steps (25 to 40 bpm)
5 bpm steps (40 to 300 bpm)
Pedi/Neo:
1 bpm steps (25 to 50 bpm)
5 bpm steps (50 to 300 bpm)