Electromagnetic emissions
Emissions Test Compliance Electromagnetic Environment - Guidance
RF Emissions
(CISPR 11)
Group 1 The NightBalance uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
RF Emissions
(CISPR 11)
Class B The NightBalance is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic Emissions
(IEC 61000-3-2)
Class A
Voltage uctuations / icker
emissions
(IEC61000-3-3)
Complies
Regulatory Information
WARNING
NightBalance should not be used adjacent to or stacked with other equipment. If
adjacent or stacked use is necessary, NightBalance should be observed to verify
normal operation in the conguration in which it will be used.
NightBalance does not have essential performance.