Test Equipment Performance Verification and Safety Tests
PageWriter TC 70 Cardiograph Service Manual 4-5
Test Equipment
The following test equipment is required to perform the complete set of Performance
Verification tests:
ECG Simulator
Electrical Safety Tester
USB memory stick with the most recent PageWriter TC70 software
Safety (S2) Equipment leakage current
(page 4-13)
SF Maximum leakage
current
<
1000 uA
Example:
S2:P/999
Safety (S3) Leads leakage current (page 4-13)
Source (Normal condition)
With Mains on applied part
(Single Fault Condition)
<
10 uA x1
<
50 uA x3
Example:
S3:P/5/20
Test Block
Name
Test or Inspection to Perform Expected (Passing)
Test Results
What to Record on a
Service Record