Troubleshooting Restarting the Cardiograph
3-34 PageWriter TC 70 Cardiograph Service Manual
Restarting the Cardiograph
To restart the cardiograph:
1
Press the On/Standby button for 2-3 seconds to shut down the cardiograph.
2 Press the On/Standby button again to power on the cardiograph.
3 If steps 1 and 2 do not work, press the restart button (rear of cardiograph). See Figure 3-1.
After approximately 20 seconds, the PageWriter TC70 software identification screen
appears, followed by an audible beep.
Figure 3-1 Restarting the Cardiograph
4
If steps 1,2, and 3 do not work, remove the batteries and the AC power from the
cardiograph.
5 Reapply power and repeat steps 1 through 3.
Restart button
(Gentle pressure only)