PAGE 9-71097626, VER. 03
b. Let the device run for at least 10 minutes on 0.5 LPM continuous flow. Record the results of
the O2 purity from the O2 Analyzer and the internal O2 sensor on the test data sheet.
41. Verify the functionality of the control panel buttons. Then turn the device off and listen for three
audible tones signifying the device turning off. Record results on the test data sheets.
42. If the SimplyGo fails testing, turn the unit completely off and retest the unit.
43. If the unit fails a second time refer to the trouble shooting section of the Service Manual.
44. Complete all the required information on the test data sheet and save the test data sheet in accor-
dance with your facility test record storage requirements.
Balancing Procedure
Perform only if a Main PCA, Sieve Bed, Compressor, or Process Valve has been replaced.
1. Set the device to a setting of 0.5 in continuous mode.
NOTE
If performing the Final Test and Balance Test due to the replacement of either the Main PCA, Sieve Bed,
Compressor, or Process Valve, perform the Balance Test.
NOTE
Power the device ON