PAGE 7-211097626, VER. 03
7.10 SIEVE BED BALANCE VALVE REPLACEMENT
Removal:
1. Remove the Rear Enclosure.
2. Remove the Front Enclosure.
3. Remove the wire harness from J14 on the Main PCA.
4. Remove the two screws that secure the Sieve Bed Balance Valve to the spine assembly.