Chapter 8
8-12 Respironics V60 Ventilator User Manual 1047358 Rev N
Alarms and messages
Check Vent: Barome-
ter Sensor Range Er-
ror
Default barometric pres-
sure of 686.0 mmHg (ap-
proximately 900 m/2953 ft
above sea level) used in cal-
culations
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(19)
Yes No No
Check Vent: Battery
Failed
Battery problem Check the patient. Connect
the ventilator to AC. Provide
alternative ventilation. Have
the ventilator serviced.
High
(35)
Yes No No
Check Vent: Battery
Temperature High
Battery problem Check the patient. Connect
the ventilator to AC. Check
for causes of overheating,
such as high room tempera-
ture, blocked vents, clogged
air inlet filter, or nonfunc-
tional fan. Provide alterna-
tive ventilation. Have the
ventilator serviced.
High
(34)
Yes No No
Check Vent: Blower
Temperature High
Technical failure Check the patient. Check for
causes of overheating, such
as high room temperature,
blocked vents, clogged air
inlet filter, or nonfunctional
fan. Provide alternative ven-
tilation. Have the ventilator
serviced.
High
(33)
Yes No No
Check Vent: Blower
Stalled
Technical failure Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High (2) No No No
Check Vent: Cooling
Fan Speed Error
Overheating of ventilator
possible
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(36)
Yes No No
Check Vent: CPU
PCBA ADC Failed
Technical failure Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(29)
Yes No No
Check Vent: Data Ac-
quisition PCBA ADC
Failed
Technical failure Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(27)
Yes No No
Check Vent: Flash
File System Error
Technical failure Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(37)
Yes No No
Table 8-3: Check Vent alarm messages: summary and troubleshooting (continued)
Message Description Corrective Action
Priority
type (ID)
Manually
resettable
Autore-
settable
Silence-
able