Chapter 9
9-2 Respironics V60 Ventilator User Manual 1047358 Rev N
Care and maintenance
Approved Cleaning Agents
Thefollowing
cleaningagentsareacceptableforuseonthetouchscreenand
exteriorsurfacesoftheventilator:
•Clean water
Soapy water with Medivators Intercept Detergent*, per manufacturer's
recommendation at 1/3 oz (10 mL) per gallon of warm tap water. Or
equivalent.
*(Benzalkonium chloride 4.8%, diethylene glycol monoethyl ether
4.8%, lactic acid 1.4%, alkyl polyglycoside 1.4%)
Cleaning Instructions
1. Apply cleaning agent to a soft lint-free cloth or use a disposable wipe.
The cloth or wipe should be saturated but not dripping.
2. Wipe cleaning agent over the entire exterior surface and touchscreen
of the ventilator.
3. Continue wiping until all visible contaminants and soiling are
removed.
4. Rinse with a clean, water-dampened cloth and allow to dry completely
before reuse.
Exterior and
Touchscreen
Disinfection
CAUTION: To prevent possible damage to the ventilator, use only those cleaning
and disinfecting agents listed in this manual.
CAUTION: To prevent possible damage to the ventilator, do not drip or spray any
liquids directly onto any surface including the front panel,
touchscreen, and navigation ring.
CAUTION: Never clean or disinfect the touchscreen with an abrasive brush or
device, since this will cause irreparable damage.
CAUTION: Do not attempt to sterilize or autoclave the ventilator.
NOTE: Use of unapproved cleaning and disinfecting agents may cause
damage to the enclosure, touchscreen, or parts of the ventilator.