Chapter 9
9-6 Respironics V60 Ventilator User Manual 1047358 Rev N
Care and maintenance
Cleaning or replacing the cooling fan filter
Clean or replace the cooling fan filter as follows, referring to Figure 9-2:
1. Insert a small, flat blade driver tip between the foam filter and the
filter retaining cover (Figure 9-2).
2. Gently pry the filter cover from the back of the ventilator. Do not
remove the fan retaining pins.
3. Wash or rinse the filter. Let it dry completely before reinstalling.
4. Replace the filter, then snap the filter cover into place.
Figure 9-2: Replacing the cooling fan filter
Filter cover
Foam filter
Fan (Do not
remove)
Fan retaining pin
(Do not remove)