Chapter 9
Care and maintenance
1047358 Rev N Respironics V60 Ventilator User Manual 9-7
Removing and replacing the battery
See “Installing the optional battery” on page A-3.
Disposal Dispose of all parts removed from the device according to your institution’s
protocol. Follow all local, state, and federal regulations with respect to
environmental protection, especially when disposing of the electronic device or
parts of it (for example, oxygen cell, batteries).
Storage between
patient use
Follow the steps below when storing the ventilator between patient use:
1. Ensure that the patient circuit is assembled and installed properly as
described in “Installing the patient circuit” on page 5-3.
2. Plug the ventilator into an AC outlet and verify that the power source
symbol is displayed.
3. Adjust settings to hospital defaults.
4. Check oxygen cylinder fill status (if applicable).
5. Ensure oxygen cylinders are turned off.
6. Ensure the environmental specifications are met:
Temperature: 5 to 40 ºC (41 to 104 ºF)
Relative humidity: 15 to 95% (noncondensing)
Service and Repairs For technical service or repair information not included in this chapter, contact
Philips.
A Respironics V60 Ventilator Service Manual is available, PN 1049766. The
Service Manual includes removal and installation procedures, parts lists, and
testing and troubleshooting information.