Chapter 10
10-4 Respironics V60 Ventilator User Manual 1047358 Rev N
Technical specifications
Alarms Table 10-4 lists the adjustable alarm ranges and resolutions. Table 8-2 on
page 8-6 describes other, nonadjustable alarms.
Menu window
settings
Table 10-4: Adjustable alarm ranges and resolutions
Parameter Range Resolution Factory default
Hi Rate (High Rate
Alarm)
5 to 90 BPM 1 BPM 30 BPM
Lo Rate (Low Rate
Alarm)
1 to 89 BPM 1 BPM 10 BPM
Hi V
T
(High Tidal Vol-
ume Alarm)
200 to 3500 mL
BTPS
5 mL 2500 mL
Lo V
T
(Low Tidal Vol-
ume Alarm)
OFF, 5 to 1500 mL
BTPS
5 mL OFF
HIP (High Inspiratory
Pressure Alarm)
5 to 50 cmH
2
O1 cmH
2
O50 cmH
2
O
LIP (Low Inspiratory
Pressure Alarm)
OFF, 1 to 40 cmH
2
O1 cmH
2
OOFF
Lo
E
(Low Minute
Ventilation Alarm)
OFF, 0.1 to
99.0 L/min BTPS
0.1 L/min OFF
LIP T (Low Inspiratory
Pressure Delay Time
Alarm)
5 to 60 sec 1 sec 20 secs
Table 10-5: Menu window settings and ranges
Parameter Range
Brightness 1 to 5
Loudness 1 to 10
Mask/ET Selection ET/Trach, 1, 2, 3, 4, Other
Exhalation Port Selection DEP (Philips Respironics Disposable Exhalation Port,
Whisper Swivel (Philips Respironics Whisper Swivel),
PEV (Philips Respironics Plateau Exhalation Valve,
Other (Other Exhalation Port), None (No inline circuit
exhalation port)
Screen Lock Off, On
Auto-Trak+ (optional) Trigger: Normal, +1 to +7.
E-Cycle: -2 to -1, Normal, +1 to +6