EasyManua.ls Logo

Philips Respironics V60 - 7.15 Power Management (PM) PCBA

Philips Respironics V60
194 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 10
10-6 Respironics V60 Ventilator User Manual 1047358 Rev N
Technical specifications
Environmental
specifications
Pneumatic
specifications
Table 10-8: Environmental specifications
Parameter Specification
Temperature Operating: 5 to 40 ºC (41 to 104 ºF)
Storage: -20 to 50 ºC (-4 to 122 ºF)
Relative humidity Operating: 15 to 95% (noncondensing)
Storage: 10 to 95% relative (noncondensing)
Barometric pressure 600 to 765 mmHg (approximately -61 to 1951 m
(-200 to 6400 ft) relative to sea level)
Table 10-9: Pneumatic specifications
Parameter Specification
High-pressure oxygen supply Connector: DISS male, DISS female, NIST
Pressure: 2.76 to 6.00 bar / 276 to 600 kPa / 40 to
87 psig
Flow: 175 SLPM
Connector: SIS
Pressure: 3.31 to 6.00 bar / 331 to 600 kPa / 48 to
87 psig
Flow: 175 SLPM
High-pressure oxygen supply
(using V60 manifold)
Connector: DISS male, DISS female, NIST
Pressure: 3.10 to 6.00 bar / 310 to 600 kPa / 45 to
87 psig
Flow: 175 SLPM
Connector: SIS
Pressure: 3.66 to 6.00 bar / 366 to 600 kPa / 53 to
87 psig
Flow: 175 SLPM
Air supply Integrated blower
Inspiratory outlet
(to patient port)
Connector: ISO 15 mm female/22 mm male conical