Chapter 10
Technical specifications
1047358 Rev N Respironics V60 Ventilator User Manual 10-7
Electrical
specifications
Accessory
Requirements
To meet performance specifications, the ventilator requires a patient circuit
and filters that meet the requirements in Table 10-11.
Other specifications
Table 10-10: Electrical specifications
Parameter Specification
AC voltage 100 to 240 VAC
AC frequency 50/60 Hz
AC power 300 VA
Battery (optional) PN 1076374: 14.4 V, 11.0 Ah, 163 Wh
Maximum system current draw: 11 A
Charge voltage: +16.9 V maximum
Operating time: 360 minutes under normal
conditions
Table 10-11: Accessory Requirements
Parameter Specification
Compliance For volume accuracy requirements, the patient circuit
compliance should be 0.98 mL/cmH
2
O.
For all other performance requirements except
volume accuracy, the maximum compliance of the
circuit can be up to 2.8 mL/cmH
2
O.
Resistance Maximum resistance of the breathing circuit and at-
tachments: 2.9 cmH
2
O at 60 L/min.
Table 10-12: Other specifications
Parameter Specification
Flow delivery 150 L/min at 40 cmH
2
O at 1951 m (6400 ft) altitude
(10% degradation in flow at 2286 m (7500 ft))
Flow range -240 to 240 L/min BTPS
Pressure range 4 to 40 cmH
2
O
Dynamic pressure regulation ± (2 cmH
2
O + 4% of target)
NOTE: Negative (subatmospheric) pressure settings
are not available.