10-8 Respironics V60 Ventilator User Manual 1047358 Rev N
Technical specifications
Start-up time Ready to ventilate 9 seconds after power on
Triggering, cycling, and leak tol-
erance
As per the Digital Auto-Trak
Sensitivity algorithms (see
“Auto-Trak Sensitivity” on page 4-3)
Inspiratory and expiratory pres-
sure drop following equipment
failure: measured at patient
connection, when the recom-
mended breathing system is in
use.
< 4 cmH
2
O (at 60 LPM)
< 1.5 cmH
2
O (at 30 LPM)
Time required for the oxygen
concentration to change from
21% to 90%
The ventilator adjusts O
2
within one breath.
FiO
2
within the gas delivery system and entire breath-
ing circuit adjusts at the following rate:
Up to 3 seconds for delivered volume of 500 mL, sin-
gle limb 22 mm ID patient circuit.
Audio alarm loudness* Highest volume setting: Average sound pressure level
is approx. 76 dB(A)
Lowest volume setting: Average sound pressure level is
approx. 62 dB(A)
Operational acoustics
*
Average sound power level is approx. 54 dB(A) mea-
sured at the ventilator
Average sound pressure level is approx. 46 dB(A) mea-
sured 1 meter from the ventilator
* In accordance with 3rd Edition testing methods
Table 10-12: Other specifications (continued)