Appendix B
Communications interface
1047358 Rev N Respironics V60 Ventilator User Manual B-13
Using Philips
IntelliBridge or
VueLink
Using Philips monitors and the IntelliBridge Open Interface
NOTE: Data displayed on the IntelliBridge system is for reference purposes
only. Decisions for patient care should not be based solely on the data
obtained through the IntelliBridge system.
The Respironics V60 Ventilator can communicate with a Philips patient
monitor using the IntelliBridge Open Interface. Figure B-3 shows the required
hardware setup. The
IntelliBridge Open Interface requires a ventilator baud rate
19,200. Check for the correct baud rate in the ventilator diagnostic mode (see
“Baud Rate” on page E-10).
Figure B-3: Connect via IntelliBridge to a Philips patient monitor
90 Unused ◆◆◆◆◆◆ N/A N/A N/A Always output as
"◆◆◆◆◆◆"
91 Occlusion alarm status
or I-time too long alarm
status
NORMAL N/A NORMAL
RESET
ALARM
N/A Report highest
urgency of these
alarms: Patient
Circuit Occluded and
Patient Disconnect
92 to 95 Unused ◆◆◆◆◆◆ N/A N/A N/A Always output as
"◆◆◆◆◆◆"
96 Parameter control setting I-TIME N/A N/A N/A Always output "I-
TIME"
97 Unused ◆◆◆◆◆◆ N/A N/A N/A Always output as
"◆◆◆◆◆◆"
98 Stop code 0x03 N/A N/A N/A ASCII End
Transmission
character (ETX)
Table B-2: SNDA record format (continued)
Field Description Example Resolution Range Units Comments
Philips patient monitor
IntelliBridge module
PN 865115 A01
IntelliBridge interface cable
PN 865114 #102/L02
EC5 module
PN 865114 L03