1047358 Rev N Respironics V60 Ventilator User Manual E-1
Appendix E. Diagnostic mode
In the diagnostic mode you select the language of software display, set the
date and time, select pressure units, enable software options, and calibrate the
touchscreen.
NOTE: The diagnostic mode is primarily for use by authorized service
personnel to download software and perform other diagnostic
procedures.
Entering the
diagnostic mode
WARNING: To prevent possible patient injury, do not enter the diagnostic mode while
a patient is connected to the ventilator. Verify that the patient is
disconnected before proceeding.
Enter the diagnostic mode as follows:
1. Make sure the patient is disconnected and the ventilator is powered
off.
2. Press and hold the Accept button on the navigation ring and turn on
the ventilator by pressing the ON/Shutdown key. The screen displays
Press again for Diagnostics or wait for Ventilation.