Appendix E
Diagnostic mode
1047358 Rev N Respironics V60 Ventilator User Manual E-11
Alarm Volume Escalation
The Alarm Volume Escalation function lets you enable or disable volume
escalation
1
. When alarm volume escalation is Enabled and a high priority alarm
is not responded to within 40 seconds, the ventilator alarm volume increases
to maximum over an 20-second period.
When the Alarm Volume Escalation function is active and a touchscreen or
button press is detected, the ventilator automatically returns the alarm volume
to the user setting.
1. From the System Settings screen, select Alarm Volume Escalation to
display the Set Alarm Volume Escalation screen (Figure E-10).
Figure E-10: Set Alarm Volume Escalation
2. The current setting is shown in the Information box at the top of the
screen. If Alarm Volume Escalation is currently Disabled, you will see a
selectable Enable button. If Alarm Volume Escalation is currently
Enabled, you will see a selectable Disable button. Press the button to
change the setting.
3. The new setting is applied after the V60 Ventilator is shut down and
powered on again.
1. Available in Revision 2.30 software and above.