Appendix E
E-14 Respironics V60 Ventilator User Manual 1047358 Rev N
Diagnostic mode
Touchscreen
calibration
Calibrate the touchscreen X and Y coordinates as follows:
1. From the Diagnostics Menu, select Touch Screen Calibration. The Touch
Screen Calibration screen is displayed (Figure E-13).
NOTE: If the Touch Screen Calibration button does not respond, press the
Accept button on the navigation ring to begin.
Figure E-13: Calibrate Touch Screen screen
2. Follow the steps shown. Press on the middle of each target with a
blunt, narrow object.
If the calibration is not successful, have the ventilator serviced.
Exiting the diagnostic
mode
Exit the diagnostic mode by turning off ventilator power with the ON/Shutdown
key.