Chapter 3
General information
1047358 Rev N Respironics V60 Ventilator User Manual 3-13
Starting up the
ventilator
NOTE: Upon power-on the ventilator automatically runs a test of the backup
audible alarm followed by the primary audible alarm. You should hear
a high-pitched tone, followed by a beep. If you do not hear all of these
sounds, discontinue use of the ventilator and have it serviced.
1. Power on the ventilator with the ON/Shutdown key.
2. Verify the ventilator operation, as described on page 5-5.
Shutting down the
ventilator
Shut down the ventilator as follows:
1. Press and release the ON/Shutdown key. The Shutdown window opens.
2. Select Ventilator Shutdown. The ventilator shuts down.
NOTE: Improper shutdown may cause a Power has been restored message the
next time the ventilator is turned on.
NOTE: If the screen is blank and the dialogue box cannot be displayed, shut
down the ventilator by pressing the ON/Shutdown key, then the Accept
button on the navigation ring.
Training Product training is available. Contact your local Philips sales representative or
Philips Customer Support for assistance. Call 1-800-225-0230 for ordering
and 1-800-722-9377 for service.