Chapter 4
Principles of operation
1047358 Rev N Respironics V60 Ventilator User Manual 4-7
Ventilation modes The Respironics V60 Ventilator operates in the following ventilation modes:
CPAP (continuous positive airway pressure) mode
S/T (spontaneous/timed) mode
PCV (pressure-controlled ventilation) mode
AVAPS (average volume-assured pressure support) mode (optional)
PPV (proportional pressure ventilation) mode (optional)
Table 4-1 summarizes the characteristics of these modes. Note that on the
ventilator, the Timed breath indicator means the breath is ventilator triggered,
while the Spont breath indicator means the breath is patient triggered.
Table 4-1: Characteristics of Respironics V60 ventilation modes
Mode
Timed breaths Spont breaths
Trigger
*
* A trigger variable starts inspiration.
Limit
A limit variable can reach and maintain a preset level before inspiration ends but it does not
end inspiration.
Cycle
A cycle variable is a measured parameter used to end inspiration.
Trigger Limit Cycle
CPAP N/A N/A N/A Auto-Trak Pressure Auto-Trak
PCV Time Pressure Time Auto-Trak Pressure Time
S/T Time Pressure Time Auto-Trak Pressure Auto-Trak
AVAPS Time Pressure Time Auto-Trak Pressure Auto-Trak
PPV Time Pressure Time Auto-Trak Pressure Auto-Trak