Chapter 6
Operation
1047358 Rev N Respironics V60 Ventilator User Manual 6-5
Using the Ramp Time
function
The Ramp Time function helps your patient adapt to ventilation by gradually
increasing inspiratory and expiratory pressure (IPAP and EPAP/CPAP) from
subtherapeutic to user-set pressures over a user-set interval. Table 6-3 on
page 6-23 describes this function’s principles of operation.
Follow these instructions to use the Ramp Time function:
1. Select the Ramp Time button in the Settings window.
The ramp starts. As the ramp progresses, the Ramp Time button
graphic fills in.
2. To change the ramp interval or to end the ramp, select the Ramp Time
button again. The Ramp in Progress window opens.
3. To end the ramp and apply the full IPAP and EPAP/CPAP immediately,
select End Ramp.
4. To end the ramp and start a new one, select Start New Ramp. The Ramp
Time setting window opens again so that you can set up a new ramp.
Using the 100% O
2
function
NOTE: The 100% O
2
feature is available in Revision 2.30 software and
above.
The 100% O
2
function delivers 100% oxygen to the patient. It is available
during Screen Lock status.
Follow these instructions to use the 100% O
2
function:
1. Select the 100% O2 button in the main GUI window.
Ramp
status
Bar