EasyManua.ls Logo

Philips Respironics V60 - Touch Screen Diagnostics

Philips Respironics V60
194 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 6
Operation
1047358 Rev N Respironics V60 Ventilator User Manual 6-7
Using PPV
Follow these instructions to set up the ventilator in the PPV mode, referring to
Figure 6-3. For principles of operation, see “PPV mode (optional)” on page 4-13.
1. Open the PPV Settings window.
2. Set EPAP, O
2
, alarm limits, and backup settings to appropriate values.
The HIP alarm limit should be greater than Max P. See “Principles of
operation” on page 4-1 for a detailed explanation of these settings.
3. Set the Max V and Max P limits.
4. Set alarm limits to appropriate values. The HIP alarm limit should be
greater than the Max P.
About Max V and Max P alarms and alarm limits
Max V (PPV maximum volume limit) and Max P (PPV maximum pressure limit)
are used to prevent the delivery of excessive pressure or volume.
WARNING: PPV limits are not intended to be the primary ventilator alarms and should
not be substituted for the alarms found in the Alarm Settings window.
WARNING: To prevent the delivery of excessive pressure or volume, set the PPV
limits appropriately. Delivery of excessive pressure or volume can occur
from a sudden increase in mask leak, inappropriate settings, or a plugged
or kinked proximal pressure line. Conversely, insufficient treatment may
result if limits are set too low.
When the Max V (PPV maximum volume limit) is reached, the breath is
terminated and a message is displayed. After the limit is reached in three
consecutive breaths, the audible alarm sounds. A PPV waveform with Max V is
shown in Figure 6-1.

Table of Contents

Other manuals for Philips Respironics V60

Related product manuals