Chapter 6
6-14 Respironics V60 Ventilator User Manual 1047358 Rev N
Operation
4. Select the desired exhalation port type (Table 6-2). Select Accept to
apply.
If you select an exhalation port that is not compatible with the
selected mask, Not allowed with current mask is displayed.
NOTE: ET/tracheostomy tubes and most Philips Respironics masks require
the use of an exhalation port. If you selected ET/Trach or Leak 1 as a
mask/patient interface, you may not select None as an exhalation port.
Table 6-1: Mask/patient interface selections
Mask/patient interface type
*
Description
ET/Trach ET or tracheostomy tube
Leak 1 Mask with minimal intentional leak characteristics. Enter
Leak 1 for any of these Philips Respironics masks:
• Contour Deluxe nasal mask
• PerformaTrak mask
• Image3 full face mask
• AF421, AF531 (EE)
Leak 2 Mask with medium intentional leak characteristics. Enter
Leak 2 for this mask:
• Philips Respironics PerforMax oro-nasal mask [EE]
• AF421, AF531 (EE)
Leak 3 AP111
Leak 4 Philips Respironics Total full face mask
Other Mask not manufactured by Philips Respironics
NOTE: If you select Other, the ventilator displays
Tot.Leak rather than Pt. Leak.
* A leak symbol is printed on Respironics masks.