EasyManua.ls Logo

Philips Respironics V60 - Running the exhalation port test

Philips Respironics V60
194 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 6
6-16 Respironics V60 Ventilator User Manual 1047358 Rev N
Operation
Running the
exhalation port test
The exhalation port test is required and its window is automatically displayed
when PEV or Other is selected.
Procedure
Run the test as follows:
1. Disconnect the patient circuit from the mask/patient interface.
2. Occlude the circuit outlet. Select Start Test.
3. Wait while the test runs.
4. Verify that Test Passed is displayed.
Progress
bar

Table of Contents

Other manuals for Philips Respironics V60

Related product manuals