Chapter 6
6-18 Respironics V60 Ventilator User Manual 1047358 Rev N
Operation
Vent Info (ventilator information)
The Ventilator Information window displays software version and other
information specific to your ventilator.
Screen Lock
Screen Lock deactivates all buttons and tabs on the touchscreen except Alarm
Silence, Alarm Reset, the Alarm/Message button, and Help. Tabs are grayed out
as in this example.
This message bar is displayed at the top of the screen:
To unlock the screen, press the Accept button in the center of the navigation
ring.
NOTE: If Screen Lock is active, the touchscreen remains locked even if an
alarm becomes active.
Auto-Trak+
The Normal Auto-Trak settings work well for most patients. Pediatric patients,
however, may benefit from more sensitive trigger settings, while some adult
patients may benefit from more or less sensitive cycle settings.
Changing Auto-Trak+ settings
1. Select Auto-Trak+ from the Menu window.