Chapter 7
7-20 V60 Ventilator Service Manual 1049766 Rev A
Component Removal/Installation
7.19 Left Side Wall Follow these steps to remove the left side wall (Figure 7-19). Reverse to
install.
1. Remove front panel (section 7.10).
2. Remove the battery (section 7.3).
3. Remove the PM PCBA (section 7.15).
4. Detach power harness assembly connector from left side wall.
5. Remove screws that attach the left side wall to the base.
NOTE: At reinstallation, install the power harness connector with its red wires
toward the top of the side wall.
Figure 7-19: Removing the Left Side Wall
M4 x 8 screw (x4)
(113 +
11.3 N cm/
10 +
1 in.-lbf)
Power
harness
assembly
Left side wall