1049766 Rev A V60 Ventilator Service Manual A-1
Appendix A Respi-Link
This section describes the following Respi-Link functions:
Downloading V60 ventilator software
Installing a V60 ventilator option
Remote troubleshooting
A.1 Downloading
Ventilator Software
Follow these steps to download V60 ventilator software using Respi-Link.
NOTE: Do not connect the null modem cable until instructed to do so.
1. Turn on the PC and verify that the Gateway interface is running by
double-clicking on Start Auto Respi-Link Gateway in the Respi-Link
folder on your desktop.
2. If the ventilator is not already in Diagnostic mode, press and hold the
nav-ring Enter button and turn on the ventilator by pressing the
On/Shutdown button on the user interface. Within 5 seconds, release
and press the nav-ring Enter button again to enter Diagnostic mode.
3. Connect a standard 9-pin male-female RS-232 null modem cable and
9-pin to 25-pin female-male adapter between the PC serial port and
the V60 ventilator.
4. Within 30 seconds, the GatewayInterface screen appears on the PC
monitor if a software package is pending (Figure A-1).