°C /°F Switchable
Temperature readings are available in Celsius or Fahrenheit (located in the upper right corner of LCD). The AccuexPro
+
® has been set
to Celsius readings by default. To switch the reading mode to Fahrenheit, press and release the On / O button. When the display shows
the last temperature taken, press and hold the On / O button until “C” appears and then press again to toggle between °C /°F modes.
Cleaning and Disinfection Instructions
Before and after each use, wipe the thermometer several times with a soft, clean cloth and 75% Ethanol or Isopropyl alcohol.
To prevent damage to the thermometer:
1. The unit must not be sterilized by boiling, gas, or steam
2. Do not immerse or soak in either alcohol or hot water
3. Do not use ultrasonic washing to clean the thermometer
4. Do not use benzene, thinner, gasoline or other strong solvents to clean the thermometer
For stubborn stains, wipe the thermometer with a clean cloth that has been dampened with water
or a neutral detergent solution. Finish by wiping with a soft dry cloth.
The use of any other solution or cleaning method beyond the above noted instructions may result in reliability issues and voids warranty.
Disposal
Disposal of this product should be carried out in accordance with local regulations for the disposal of electronic products.
Product Specications
This device conforms to the following standards: IEC 60601-1-11:2015 Medical electrical equipment — Part 1-11: General requirements for basic safety and essential performance —
Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment; IEC 60601-1-2:2014 Medical electrical equipment -
Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests; EN 60601-1:2006 Medical electrical
equipment. General requirements for basic safety and essential performance (+A12:2014); ANSI/AAMI ES60601-1:2005 Medical Electrical Equipment, Part 1: General Requirements For Basic Safety
And Essential Performance. The manufacturer is ISO 13485 certied.
Limited Warranty
The Physio Logic® AccuexPro
+
® Thermometer is guaranteed for one year from the date of purchase. If the thermometer does not function properly due to defective components or poor workman-
ship, we will repair or replace it free of charge. All components are covered by this warranty excluding the battery. The warranty does not cover damages due to improper handling of this thermometer.
To obtain warranty service, an original or copy of the sales receipt from the original retailer is required.
Type Digital Thermometer
Measure Range: 32 °C – 42.9°C (90°F-109.2°F)
Accuracy:
±0.1°C (±0.2 °F) during 35.5°C – 42.0°C (95.9 °F – 107.6 °F)
at 18°C–28°C (64.4 °F – 82.4 °F) ambient operating range
±0.2°C (±0.4 °F) for other measuring and ambient operating range
Operating Mode: Predictive Mode: Oral, Rectal, Underarm – Direct Mode: Bath
Display: Liquid crystal display, 3 1/2 digits
Memory: Last Measurement
Battery: One 3.0 V DC Button Battery Size CR2032
Battery Life: Approximately 200 hours of continuous operation or 1 year with 3 measurements per day
Dimension: 13.4 cm x 3.1 cm x 1.6 cm (L x W x H)
Weight: Approximately 23 grams including battery
Expected Service Life: Three years
Ambient Operating Range:
Temperature: 5°C – 40°C (41°F –104°F)
Ambient Operating Range: Relative Humidity: 15%–95%RH
Atmospheric Pressure: 700hPa – 1060hPa
Storage and
Transportation Condition:
Temperature: -20°C–55°C (-4°F–131°F)
Relative humidity: 15% – 95%RH
Atmospheric Pressure: 700hPa - 1060hPa
Ingress Protection Rating: IP 27
Classication: Type BF
General Care
1. Clean the thermometer before and after each use.
5. Avoid dropping the thermometer or subjecting
it to strong shock.
2. Store the unit in its protective case when it is not in use. 6. Do not attempt to disassemble the unit.
3. Do not bite the probe.
7. The thermometer should only be used under
the supervision of an adult.
4. Do not store the unit where it will be exposed to direct
sunlight, dust or humidity. Avoid extreme temperatures.
8. Do not walk, run, or talk during temperature taking.
Direct Current
Consult Accompanying Documents
Batch Code
Storage and Transportation
Temperature Limit: -20 °C – 55 °C (-4 °F – 131 °F)
ManufacturerType BF Applied Part
Troubleshooting
Error Message Problem Solution
Temperature reading is lower
than 32° C (90° F)
Turn o thermometer, wait one minute and take a new reading
following User Guide.
Temperature taken is higher
than 42.9 °C (109.2° F)
Turn o thermometer, wait one minute and take a new reading
following User Guide.
Thermometer moved while
taking temperature
Retest temperature. Stay still when taking reading.
Measurement time
exceeds 10 seconds
The probe tip may not have cooled down,
please wait 30 seconds between each measurement.
The thermometer moved while taking temperature.
Wait 30 seconds and retest temperature. Stay still when taking reading.
Drinking hot or cold uids, exercising, smoking or performing other
activities may raise or lower your temperature. It is important to relax
for approximately 5 minutes, mouth closed, prior to taking another reading.
The system is not
functioning properly
Contact the AMG Customer Service Team at: 1-800-363-2381