33
13. Once the treatment is over position the switch to 0 and remove the plug from
the mains socket; Before opening the valve and emptying the boiler wait until
the appliance is completely cooled. Then, once the bottle has been removed,
remove the accessories (separate the sprayer nebuliser cup and the underwa-
ter tube), unscrew the safety cap and empty the residual water from the boiler.
14. WARNING: The quantity of water introduced into the boiler (800cc) allows only
one inhalation to be carried out. The quantity of depleted water (collected in
the decanter) has no more therapeutic value and must therefore not be reused.
15. Proceed with the cleaning operations as indicated in the CLEANING section.
16. Check the water level inside the boiler at every treatment.
During operation of the device high temperatures are registered on
the steam conveying nebuliser cup because it is close to the aqueous steam
emission point needed for treatment and for mixture with the thermal water.
In any case this temperature does not constitute a risk as the treatment must
be carried out maintaining a distance of about 20-40cm: therefore, at this dis-
tance these temperatures are not dangerous.
Never come into contact with the steam conveying nebuliser cup during
operation of the device.
DO NOT OPERATE THE APPLIANCE WITHOUT WATER IN THE BOILER
AS THIS WILL TRIGGER A THERMAL BREAKER THAT MUST BE RESET MANU-
ALLY.
THIS THERMOSTAT MUST BE RESET AT AN AUTHORISED CENTRE AND / OR
THE PIC TECHNICAL SERVICE CENTRE.
WARNING: The power cable is the separating element from the electrical mains;
even though the appliance has a specific on / off switch the power plug must be
kept accessible when the appliance is in use in order to allow a further method if
necessary for disconnection from the mains.
THERMAL BREAKER: The appliance has a thermal breaker that trips if excessively
high temperatures are reached. If this occurs, switch off the appliance by pressing
the luminous switch (toward the 0 position) and allow it to cool for about 30 minutes.
The thermal breaker may trip for one of the following reasons:
- the appliance operated outside of the operating limits indicated in this manual.