SECTION
2.
OPERATION
(Cont.)
2.4
Measuring
Fie
l d
Strength
(Cont.)
The
field
meter
should always
be
operated
in
a
vertical
position
when making
field
strength
measurements.
After
calibrating
as
described
in
the
previous
section,
proceed
as
follows:
2.5
1.
Set
the
FULL
SCALE
switch
to
the
range
approxi.mating
the
signal
strength
e:xpected.
2.
With
the
FUNCTION
switch
set
at
FI,
rotate
the
unit
to
orient
the
loop
antenna
and
obtain
maxi.mum
deflection
on
the
panel
meter,
changing
the
position
of
the
FULL
SCALE
switch
if
nec
-
essary
to
keep
the
panel
meter
indication
on
scale.
3.
Read
the
field
intensity
directly
from
the
panel
meter,
using
the
position
of
the
FULL
SCALE
switch
as
a
guide.
For
example,
with
the
FULL
SCALE
switch
on 100
mv,
a
full
scale
reading
of
10
on
the
m
eter
means 100
millivolts
per
meter,
and a
reading
of
5.6
indicates
a
field
strength
of
56
millivolts
per
meter.
4.
Multiply
the
reading
obtained
in
step
3
by
the
K
factor
for
the
operating
frequency
given
on
the
calibration
certificate
for
the
instrument.
This
procedure
is
followed
when
it
is
necessary
to
make
full
use
of
the
specif
i ed
accuracy
of
the
field
meter.
5.
To
de
-
energize
the
field
meter,
rotate
the
FUNCTION
switch
to
the
OFF
position.
This
disconnects
the
batteries
from
the
circuit.
Closing
the
cover
also
removes
the
battery
power.
Measuring Harmonic
Ratio
in
Decibels
Field
Strength
Ratios
may
be
measured
directly
in
decibels
by
reading
the
db
scale
of
the
meter
in
conjunction
with
the
FULL-SCALE
range
indicator.
Since
each
step
of
the
FULL--SCALE
switch
corresponds
to
one (1)
decade
or
twenty
(20)
db,
the
ratio
is
determined
by
the
following
formula:
db1 - db2 + N x
20
=
Ratio
in
db
where:
db1 = Harmonic
field
strength
db2 = Fundamental
field
strength
N =
Number
of
FULL
SCALE
steps
between f\mdamental
and
harmonic