5.7. Compound Measurement Surface Velocity
Select “Compound Measurements”
Select "Surface Velocity"
Select parameter setting “b”.
Measure and enter the distance “b”
T
ransducers in first position
Start
“t1” is measured. Stable reading
indicated by a tick
Receiver to 2nd position - Start
“t2” is measured. Stable reading is
indicated by a tick and result screen
is displayed.
The display shows:
- “t1”
- “t2”
- Surface velocity
Save the result or press start to