The Pundit Lab should be zeroed
using the calibration rod on a
regular basis and in particular if the
transducer frequency is changed or if
the cables are changed. The expected
calibration value (µs) is marked on the
Couple the transducers to the
calibration rod by applying couplant
to the transducers and both ends of
the rod and pressing firmly together
When zeroing the exponential transducers the dedicated calibration rod
(325 40 174) should be used.
(See chapter 3 for more info)
Check Tx/Rx matches transducer
Correction factor set to 1.0.
If necessary, select to enter the transit
time as marked on the calibration rod.
returns to the previous menu.
returns to the previous menu.
Press “Start” to begin the calibration
The final display shows the expected
transit time and below it the measured
transit time. This should match the
value on the calibration rod.