Revision O 07/18/2017
www.pronktech.com 800-541-9802 Page 13
PRONK TECHNOLOGIES
Accessories
ECG is Noisy
The noise may point to a ground loop issue. Check for
ground loop noise by running the SimCube on battery
power to see if the noise still exists. If it is present, check to
see that the monitor under test has its A/C filter(s) enabled.
Please note that monitors may have more than one filter
that filter out of different frequencies.
ADAPT-FP
Welch Allyn Flexiport style: A snap connector, used on any monitor
with Welch Allyn Flexiport cuff.