EasyManua.ls Logo

Prostat PFM-711B - Page 21

Prostat PFM-711B
26 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
20
b. To charge the Wand to >-1.0kV, press and hold the Negative (-) button switch
8. Touch the wand to the CPM-720B’s sensing (upper) plate and note the voltage indicated on
the PFM-711B.
9. Once the desired voltage level is achieved, release the button switch and remove the PCS-
730B Charger from the air path.
10. Observe the voltage drop indicated on the PFM-711B meter. Voltage should decay over time
to the initial off set point indicated during the balance test procedure outlined above.
Body Voltage Generation
Using the PFM-711B/CPM-720B combination to Measure Body Voltage Generation related to
Footwear and Flooring combinations.
1. Mount the CPM-720B Charge Plate Assembly on the PFM-711B Field Meter Case as described
above.
2. Ground the case of the PFM-711B, as described below:
a. Using a PWS-610M fabric wrist strap cuff, mount the cuff on one of the five (5) pound
NFPA electrodes such as the PRS-801W.
NOTE
The use of the Prostat PWS-610M is recommended for this application due to its built-in 10mm
male snap so that connects to the grounding female snap on the back of the PFM-711B meter.
b. Remove the insulated cap covering the 10 mm male snap located in the cuff fabric
approximately 1.5 inches from the cuff buckle.
c. Snap mount the PFM-711B’s 10 mm female snap (located on the back of the PFM-
711B’s case) onto the wrist strap 10 mm male snap.
d. Adjust the wrist strap cuff so that the instrument is held firmly to the electrode; lock
the adjustment buckle.

Related product manuals