Guidance and manufacturer´s declaration electromagnetic immunity
Boaray 5000 C/D series is intended for use in the electromagnetic environment specified
below. The customer or the user of the Boaray 5000 C/D series should assure that it is
used in such an environment.
Immunity test
IEC 60601 test
level
Complian
ce level
Electromagnetic environment -
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Veffective
value
150 kHz80
MHz
3V/m
80 MHz2.5
GHz
3V
effectiv
e value
3V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the Boaray 5000 C/D series,
including cables, than the recommended
separation
distance calculated from the
equation applicable to the frequency
of the
transmitter.
Recommended separation distance
d =
P2.1
d =
P2.1
80 MHz800 MHz
d =
P
3
.2
800 MHz2.5 GHz
where p is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in metres
(m).
b
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a should be less than the compliance level in
each frequency range. b
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic is affected by absorption and