31
support.R82.org
Service information
Maintenance of a medical device is wholly the responsibility of the owner of that device. Failure to
maintain a device in accordance with the instructions may invalidate the device’s warranty. Furthermore,
failure to maintain a device may compromise the clinical condition or safety of users and/or their
caregivers.
1st Day
Read the User Guide carefully before
use and store it in a safe place for future
reference
Store the enclosed tools on/at the product
Daily
Use a dry cloth to clean the product
Visual inspection. Repair or replace any
damaged or worn-out parts
Ensure that all fasteners are in place and
tted correctly
Check all ttings and buckles for signs of
worn out parts
Be aware of any MAX mark indications
before adjustment
Weekly
Check that all castors are moving freely
and that all wheel locks/brakes are easily
activated
Wipe off any residue and dirt from the
product using a cloth with warm water and a
mild detergent/soap and dry before use
Check that the angle mechanism functions
correctly and that the safety lock is working
correctly
Disinfect the product. please check the
disinfection section in this User Guide for
more information
Monthly
All nuts and bolts on this product must be
checked and tightened to avoid unnecessary
failures
Oil the swing-away parts. We recommend
you use a professional lubricant system
Annually
Inspect the frame for cracks or signs for
worn-out parts and carry out annual service.
Never use a product which appears to be
substandard or faulty
Warewashing machine
This product can be washed at:
60°C in 10 minutes
...using a mild detergent in a washing machine
designed for medical devices. Use the machine’s
drying function to dry the product. Clean the
accessories separately.
Before washing
Remove the following items and clean them
separately:
Accessories
...using a mild detergent in a washing machine
designed for medical devices. Use the machine’s
drying function to dry the product. Clean the
accessories separately.
Disinfection
The product without accessories can be
disinfected with a 70% disinfectant IPA solution.
It is recommended to wipe off any residue
and dirt from the product, using a cloth with
warm water and a mild detergent/soap without
chlorine and let it dry before disinfection.
Service interval
A detailed inspection of the product must be
made every 12 months (every 6 months for heavy
users) and every time the product is reissued for
use. The inspection should be carried out by a
technically competent person who has been
trained in the use of the product.
EN