Display test proceeds (all symbols are backlit for a moment), program
name and number is displayed first, ZERO indication with reading unit
next (displayed reading unit is conditioned by scale type).
In case the weighing result is not zero after indication stabilisation, press
key.
7.4. Warm-Up Time
For correct scale operation, the workroom temperature must range
between +15C +30C.
It takes 30 minutes for the device to warm up.
During the thermal stabilization, the indications on the screen can change.
If you want to carry out adjustment operation, make sure that your device
is thermally stable.
Temperature and humidity variations during operation may cause
indication errors. To remove the indication errors perform user adjustment.
7.5. Battery Charge Status
The scale of standard design is equipped with an internal battery. Battery state
is signalled by pictogram, the pictogram is displayed in the top bar of the
display.
pictogram display mode
Meaning
No pictogram
Battery charged. Regular scale operation.
Pictogram displayed continuously
Too low battery charge (the scale is about to shut
down). Charge the battery immediately.
Blinking pictogram, blink frequency:
ca. 1 s
Battery charge in progress. The device is connected
to the power supply charging the battery.
Blinking pictogram, blink frequency:
ca. 0.5 s
Battery error. Battery is damaged.
7.6. Battery Charge Status Check
Press and keys combination.
Depending on the battery state, a respective status is displayed on the
screen for 2s:
80%
Battery power supply. Battery power given in %.
CHArGE
Battery charge in progress. The device is connected to the power supply
charging the battery.