EasyManua.ls Logo

RADWAG WTC 200 - 13 Error Messages

RADWAG WTC 200
17 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
16
No printout on a
scale-connected
printer.
Incorrect printer port set in
<5.2.1.Prt> parameter.
Enter < P5.ducE / 5.2.Prtr> submenu and
set correct <5.2.1.Prt> parameter value.
Incorrect transmission
parameters for the selected
printer port.
Enter <P4.Conn> menu and set correct
transmission parameters for the selected
printer port.
No variable declared in
weighing printout project.
Enter <P6.Prnt / 6.2.GLP> submenu and
declare variables that are to be printed.
Communication with
the additional display
not established..
Incorrect additional display
port set in <5.3.1.Prt>
parameter.
Enter < P5.ducE / 5.3.AdSP> submenu
and set correct <5.3.1.Prt> parameter
value.
Incorrect transmission
parameters for the selected
computer port.
Enter <P4.Conn> menu and set correct
transmission parameters for the selected
additional display port.
Displayed mass unit
does not comply with
the scale data plate.
Changed scale start unit in
<9.1.UnSt> parameter.
Enter <P9.Unit / 9.1.UnSt> submenu and
set unit complying with the scale data
plate.
Changed custom unit in
<9.2.Unin> parameter.
Enter <P9.Unit / 9.2.Unin> submenu and
set unit complying with the scale data
plate.
13. ERROR MESSAGES
- E r r 2 -
Value beyond zero range.
- E r r 3 -
Value beyond tare range.
- E r r 4 -
Adjustment weight or start mass out of range (1% for adjustment weight, 10 for
start mass).
- E r r 5 -
Battery error. Battery is damaged.
- E r r 8 -
Time of the following operations exceeded: taring, zeroing, start mass
determining, adjustment process.
- n u l l -
Zero value from converter.
- F U L L -
Weighing range exceeded.
- L H -
Start mass error, indication out of range (10% of start mass).
- H i -
Display range of total mass on scale display exceeded in 'Totalizing' mode.
- u L o -
Too low battery charge. The scale is about to shut down.
- E r r L o -
- Determined mass of single part in 'Parts counting' mode too small.
- Value of 'Min' threshold is greater than value of 'Max' threshold in '+/- control'
mode.
- E r r H i -
- Entered value of single part greater than maximum capacity in 'Parts counting'
working mode.
- Entered value of 'Max' threshold greater than maximum capacity in '+/- control'
mode.
- Entered reference mass greater than maximum capacity in 'Percent weighing'
mode.

Other manuals for RADWAG WTC 200

Related product manuals