- 52 -
27.6. Continuous Transmission
For continuous transmission mode the scale provides option of mass
measurement printout in basic unit and in additional unit. The mode can be
activated with command sent via computer, or by setting respective parameter
values on the scale.
Format of frame sent when <5.1.2.Cnt> parameter is set to CntA value:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S I space
stability
marker
space character mass space unit CR LF
Stability marker
[space] if measurement result stable.
[?] if measurement result unstable.
[^] if high limit is out of range.
[v] if low limit is out of range.
Character
[space] for positive values.
[-] for negative values.
Mass
9 characters with decimal point, right justification.
Unit
3 characters, left justification.
Command
3 characters, left justification.
Format of frame sent when <5.1.2.Cnt> parameter is set to Cntb value:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S U I
stability
marker
space character mass space unit CR LF
Stability marker
[space] if measurement result stable.
[?] if measurement result unstable.
[^] if high limit is out of range.
[v] if low limit is out of range.
Character
[space] for positive values.
[-] for negative values.
Mass
9 characters with decimal point, right justification.
Unit
3 characters, left justification.
Command
3 characters, left justification.