SENSe Subsystem:FREQuency EM510
5.75 4065.7763.32-01.00
. . :SPAN<numeric_value>|UP|DOWN |MINimum|MAXimum
Selection of frequency range with option IF-panorama.
The following ranges are available: 10 kHz, 25 kHz, 50 kHz, 100 kHz, 150 kHz, 256 kHz, 300 kHz, 400
kHz, 600 kHz, 800 kHz, 1200 kHz , 2400 kHz, 4800 kHz und 9600 kHz. The entered frequency must
exactly correspond to the upper values. It is not rounded.
Note:
The IF-panorama is available if the software option EM510IM (ITU Measurement) or the software option
EM510SU (Spectrum Unit) is installed. Otherwise the error message 1,"Device dependent error;Not
installed" is output.
Parameters:
<numeric_value> frequency range
UP|DOWN taking the range after|before the current bandwidth
MINimum|MAXimum sets the minimum|maximum frequency range
*RST state:
100 kHz
Example:
FREQuency:SPAN 25 kHz
. . :SPAN? [MINimum|MAXimum]
Query about the frequency range with option IF-panorama.
Parameters:
none query about the current frequency range
MINimum|MAXimum query about the minimum|maximum frequency range
Result:
Frequency value Hz
Example:
FREQuency:SPAN? 25000