EM510 SENSe Subsystem:FREQuency
4065.7763.32-01.00 5.76
. . :STARt <numeric_value>|MINimum|MAXimum
Setting the start frequency of a frequency scan.
Parameters:
<numeric_value> frequency
MINimum|MAXimum sets the lowest|highest start frequency
*RST state:
1 MHz
Example:
FREQuency:STARt 1.5 MHz
. . :STARt? [MINimum|MAXimum]
Query about the start frequency of a frequency scan.
Parameters:
none query about the current start frequency
MINimum|MAXimum query about the lowest|highest start frequency
Result:
Frequency in Hz
Example:
FREQuency:STARt? -> 1500000
. . :STEP
. . . [:INCRement] <numeric_value>|MINimum|MAXimum
Setting the stepwidth for the command SENSe:FREQuency[:CW|FIXed] UP|DOWN.
Parameters:
<numeric_value> frequency stepwidth
MINimum|MAXimum sets the smallest|largest stepwidth
*RST state:
1 kHz
Example:
FREQuency:STEP 2 kHz